High Temperature Smart Calibration Salt Bath

DTW-EM High Temperature Smart Calibration Salt Bath with Elec. Meas.

2026-10-6 This article is read 49 Times

EM- SaltBath.jpg


◇ Introduction ◇ Function◇ Parameter◇ Product order◇ Download◇ Related products

Introduction

The DTW Series High-Temperature Precision Salt Bath provides a precise, non-pressurized heat source with a non-toxic and non-corrosive working medium, making it ideal for high-precision testing of thermocouples and RTDs. The DTW-EM with Electrical Measurement Series combine temperature calibration with precision electrical testing. Featuring integrated Source and Measure capabilities, it effectively handles sensor calibration, temperature comparison, and electrical signal detection, achieving true multi-purpose operation.

Product Characteristics

Sensor Full Lifecycle
Management
     Comprehensive sensor management:add, delete, edit, view, and copy.
     Auto-loads formula coefficients for SPRT and RTC sensors based on model and range.
     Smart sensors: plug-and-play operation,automatic inventory tracking,and clear visual identification.
Module
Categories
Electrical Measurement Module 1
Supports
Standard
Type
Secondary SPRT(Pt25, Pt100)
 RTD, SMP, and other smart sensors
Supported DUT Types
Supports
Tested Type
RTDITS90 (Cu50, Cu100, Pt50, Pt100, PtX),
CVD, IEC60751, Nickel, SMP
RTCS, B, R
TCK, N, T, J, E, R, S, B, Wre3, Wre5
Transmitter
DCV, DCI
Temp. SwitchDry Contact


Integrated
Core
Hardware
Technology
Built-in
high-precision
electrical
measurement unit

Standard 6½-digit and 7½-digit modules (optional accuracy grade);

Measures resistance (Ω), DC voltage (DCV), and DC current (DCI);

Ultra-high resolution down to 0.1 μΩ, ideal for high-precision metrology tasks.

The temperature field and electrical measurement units are deeply integrated at both hardware and algorithmic levels.

4-Wire RTD
Constant-Current
Reversal
Measurement
Technology
Separates constant-current excitation from high-impedance voltage sensing circuits. Eliminates lead and contact resistance errors for exceptional measurement.
Cold Junction
Compensation
Design
Dedicated high-precision cold-junction sensor with multi-point monitoring and intelligent compensation, ensuring highly accurate thermocouple cold-junction measurements.
Automated
Calibration
Tasks
Indication
Calibration

Calibration Items: RTD, RTC, TC, DCV/DCI Transmitter, SWNTC, and other DUTs.

When paired with an electronic scanner switch, it enables fully automated multi-DUT calibration for temperature points, repeatability, and deviation—with unlimited temperature points. Raw measurement records are generated automatically.

Switching signal
Calibration

Supports temperature switches, oil-level/winding temperature controllers, and pressure-actuated thermometers. Supports simultaneous parallel verification of up to 6 channels. Automatically generates heating/cooling sequences to minimize stabilization time, and supports both ascending and descending ramp testing.

Task Configuration: Supports configuration of parameters such as ramp rate, temperature point intervals, accuracy class, and ascending/descending test parameters to complete the calibration point setup for switching signals.

Batch
Processing &
Customized
Automation
Workflows
Batch Import &
Auto-Calibration

Enables batch entry and editing of DUT serial numbers, with template-saving capability.

Supports configuration of temperature fluctuation, temperature deviation, and temperature change monitoring thresholds, used for automatic correction and pass/fail determination.

Multi-point self-calibration: Configure multiple temperature points at once and start the automated workflow with a single click.

Custom Test
Sequences
Supports programmable multi-point test sequences that automatically switch measurement modes to execute batch verification, calibration, comparison, and pass/fail determination—fully hands-free.
Intelligent Programmable
Multi-Temperature Segment
Verification Mode
Two operating modes available as needed:
Automatic programming mode: One-click setting of the full-process temperature sequence; the device executes automatically and sequentially.
Supports customized delayed start and stabilization hold time.
Multi-cycle configuration, automatic control stop upon task completion, unlimited number of temperature points, unattended completion of the entire test and verification process.
Manual programming mode: Import temperature sequences for automatic heating and stabilization, supports manual switching of temperature points as needed;
Simultaneously supports delayed start, cycle count, and shutdown settings, with an unlimited number of temperature points, facilitating on-site manual review and record comparison/marking.
Adaptive Smart
Temperature Control
Dynamically adapts gains to thermal load and medium changes. Selects the best strategy to ensure temperature uniformity and stability.
Multi-channel
Synchronous
Automatic
Acquisition
Once the temperature field stabilizes, the system automatically scans all channels in sequence and synchronously acquires data from both the temperature field and the DUT. It performs multiple ascending/descending readings as required by the regulations and automatically calculates the average, ensuring the objectivity and accuracy of the measurement data.
Regulation-Driven
Automated Data
Processing

Embeds calculation algorithms from national metrological verification regulations (e.g., JJG 229, JJG 141, JJG 351, JJG 630). Built-in intelligent stability algorithms continuously monitor the temperature change rate and automatically calculate key parameters—including indication error, hysteresis error, resistance, and temperature—eliminating manual computation.

Calibration items can be automatically matched with corresponding tolerance thresholds based on the instrument's accuracy class, outputting regulator-compliant calibration results.

Supports historical calibration record retrieval, allowing users to access raw data and full project configurations from any past record, with support for exporting raw data sheets.

Real-time
Monitoring
Throughout the
Process

Displays real-time, all-dimensional information including electrical measurement data, temperature field status, operating status, task progress, elapsed time, and stability countdown. In Indication Calibration mode, it simultaneously displays the current temperature point and the current calibration count.

Real-time deviation/tolerance calculation, with automatic highlighting of out-of-tolerance data.

Monitors the 10-minute fluctuation, maximum, and minimum values of both the standard and DUT channels in real time, with line charts provided for data visualization.

System
Management
& Data
Security
Access ControlSupports tiered access control across three categories: calibration operations, backup and restore, and advanced configuration. Fully compliant with metrology laboratory requirements.
Multiple Raw
Record Export
Options
Export via USB drive, QR code download, or other formats.
System LoggingComprehensive recording of all verification tasks, equipment operations, and fault/exception events. Automatic logging of operational data for remote diagnostics, enabling fast fault pinpointing and resolution.
Backup & Restore
Management

Restore from historical calibration records. Configurable retention period (up to 1+ year); default: 7 working days.

Supports data recovery at the individual calibration record level. Once the target record is retrieved, the system can restore the calibration configuration and data state to that specific point in time.

Resume After
Power Failure
Equipped with power-loss resume capability. In the event of a power failure, the system automatically resumes the calibration task, retaining previously measured data to prevent duplicate efforts.
Human-
Machine
Interaction
(HMI) &
Extensibility
Touchscreen
Human-Machine
Interface
5- or 7-inch HD touchscreen & graphical UI. Intuitive operation, minimal operator training.
Mass Data Storage
& Retrieval
8 GB built-in storage for 100,000+ calibration records. Advanced combined search (ID, date, project ID, instrument type, calibration date, etc.).
Communication
Interfaces
Standard interfaces: RS232/RS485, Wi-Fi, and LAN. Supports USB data export, ensuring long-term compliance and serviceability.
Sustainable
Upgrades
In-system upgrade via USB drive for the latest features and regulation revisions. Ensures long-term compliance and serviceability.

Product Selection Guide

ModelDTWDTW-EMFeatures
Rapid Salt Melting••Features an integrated rapid melting control mechanism. The medium fully melts in ≤ 2 hours, significantly reducing startup preparation time and ensuring efficient continuous calibration.
Scheduled
Auto-Melting
••Features a programmable timer for auto start/stop. Users can pre-schedule the salt melting process to maximize efficiency.
Touchscreen
Human-Machine
Interface
••7-inch HD touchscreen & graphical UI Intuitive operation, minimal operator training
Multi-Point
Self-Correction
••Multi-point self-calibrator: Configure multiple temperature points at once and start the automated workflow with a single click.
Intelligent
Programming Setup
••Features multi-point temperature programming, allowing users to preset calibration points into the device. The system executes the sequence sequentially with a one-click start, eliminating the need for repetitive parameter input.
Adjustable Ramp
Rate
••Adjustable Ramp Rates for Customized Heating and Cooling
Adaptive Smart
Temperature Control
••Dynamically adapts gains to thermal load and medium changes.Selects the best strategy to ensure temperature uniformity and stability.
Comprehensive
Sensor
Compatibility

•The system features an independent reference channel, while the DUT channels accommodate thermocouples, RTDs, transmitters, and temperature switches.
Sensor Full
Lifecycle
Management

•Comprehensive sensor management:add, delete, edit, view, and copy.
High-precision
electrical
measurement unit

•Standard 6½-digit and 7½-digit modules (optional accuracy grades);Ultra-high resolution down to 0.1 μΩ
Automated Data
Processing

•Calibration items can be automatically matched with corresponding tolerance thresholds based on the instrument's accuracy class, outputting regulation-compliant calibration results.
Multi-channel
Automatic
Acquisition

•The system automatically scans all channels in sequence and synchronously acquires data from both the temperature field and the DUT.
Real-time
Monitoring
Throughout the
Process

•Displays real-time, all-dimensional information including electrical measurement data, temperature field status, operating status, task progress, elapsed time, and stability countdown.
Access Control
•Supports tiered access control across three categories: calibration operations, backup and restore, and advanced configuration.

Technical parameters

Measurement
Channel
Measuring RangeAccuracyResolution
SPRT
Measurement
0Ω to 400Ω±0.002Ω at (0 to 50)Ω  ;
±40ppm RD at (50 to 400)Ω
0.001mΩ
RTD
Measurement
0Ω to 400Ω±0.002Ω at (0 to 25)Ω  ;
±80ppm RD at (25 to 400)Ω
0.001mΩ
TC Measurement0mv to 100mv±(0.02% RD + 5μV)0.001uV
DCI Measurement0mA to 20mA±(0.02% RD + 2μA)1nA
DCV
Measurement
0V to 10V±(0.02%RD + 2mV)1uV
ModelDTW-EM560GDTW-560GDTW-EM670GDTW-670GDTW-EM850GDTW-850G
Integrated
Electrical
Measurement
•••
Temp. range (℃)180 to 560450 to 670500 to 850
Stability
(℃/10min)
±0.01±0.01±0.01
Radical
Uniformity
≤0.01≤0.01≤0.01
Axial Uniformity≤0.02≤0.02≤0.02
Access Opening ×
Depth (mm)
φ150×400φ150×400φ150×400
Rated Power3kW3kW3kW
Rapid Salt
Melting
Features an integrated rapid melting control mechanism. The medium fully melts in ≤ 2 hours, reducing startup preparation time and ensuring efficient continuous calibration
Scheduled
Auto-Melting
Features a programmable timer for auto start/stop. Users can pre-schedule the salt melting process to maximize efficiency.
NoteCustomizable 1100 ℃ liquid bath for special media.
DTW-EM High Temperature Smart Calibration Salt Bath with Elec. Meas.
  • Online message

  • Your name
    Your title must be filled
  • Cellphone number
    Mobile phone number is required
  • whatsapp/wechat
    Optional
  • Type of consultation
    Required, please select the type of consultation
  • Consultation content
    Required, what is the content you want to consult?
  • Verification code: Please fill in the calculation result.
Previous:DTS-EM100BG Ultra-Low Temperature Smart Calibration Bath with Elec. Meas.next:DTS-CT10G Smart Refrigerated Calibration Bath (-10℃ to 105℃)

Copyright© Tai'an Dearto Automation Instrument Co., Ltd.

Chinese